: Keep the sensor area clean and dry. For the ion balance plate, clean ultrasonically if dirt or condensation interferes with readings.
Calibrates the meter to zero. It removes residual charges on the sensor before taking a new reading. LED Focus Rings (Distance Guides)
If your unit has an Ion Balance plate (a small metal plate often stored on the back), attach it to the sensor end and press the IB button to measure the balance of your ionizers. Maintenance & Safety
Press the HOLD button to freeze a measurement on the LCD display, which is helpful when the screen is difficult to see during the test. 3. Ion Balance Mode simco fmx002 manual full
, acting as a guide for its full operation, troubleshooting, and maintenance. 1. Product Overview and Key Features Simco-Ion FMX-002
A flat, polished aluminum plate (typically 6" x 6") mounted on an insulated stand.
: Exactly 1 inch ± 1/64 inch (25 mm ± 0.5 mm) from the targeted surface. Voltage Measuring Range : Low Range : 0 to ±1.49 kV High Range : ±1.0 kV to ±20.0 kV Ion Balance Mode Range : 0 to ±200 V System Accuracy : ±10% of the reading Response Time : Less than 1 second updates : Keep the sensor area clean and dry
Like all precision measurement tools, the FMX-002 requires periodic recalibration to ensure its accuracy within the ±10% specification. The original manual strongly advises against any user attempts at disassembly, modification, or calibration. SIMCO's official guidance states that to ensure data accuracy, the unit should be returned to an authorized service center for calibration after one year of use.
The FMX-002 provides visual and audible indicators to help the user identify operational issues:
The FMX-002 features a four-button interface and a multi-color hybrid LCD that displays both numeric measurements and real-time trends. Control Panel Buttons FMX-004 - Simco-Ion It removes residual charges on the sensor before
This comprehensive technical guide serves as an operational manual, troubleshooting resource, and calibration reference for the Simco FMX-002. 1. Technical Specifications and Capabilities
SIMCO FMX-002 Electrostatic Fieldmeter: Complete Operation Manual